Thank you very much for purchasing the SUPERIOREX II ODS column for HPLC.
SUPERIOREXII ODS is a HPLC column filled with modified ODS silicagel particle based on precision-classed, high-quality spherical silicagel.
In order to use SUPERIOREXII
ODS for a long period of time and stably, please read this instruction manual carefully, and use it correctly.
See Fig. 1 for the column connection. If the tubing is inappropriate, especially if a tube for a different type of column is used, the length after the ferrule tip (V in Fig. 1) is often different from the end fitting length L, and a problem
may occur.
* For the end fitting of each column, check the column index (Specification list).
(L>V) If L is greater than V, dead volume may be generated and cause peak broadening or tailing or deterioration of separation performance.
(V>L)
If L is smaller than V, liquid leak may occur because of inadequate ferrule adhesion.
Therefore, we recommend replacing the ferrule together with the column.
* If the column is replaced frequently, the male nut may crush the ferrule
and liquid may leak. Since tightening the nut too much may cause its head to come off, replace the ferrule at an early stage.
Problems in high performance liquid chromatography are attributable to various causes that cannot all be enumerated. The table below describes some comparatively common problems related to the column.
Symptom | Cause | Measures |
---|---|---|
1. Column pressure rise. | Blocking with foreign matter 1. Dust or insoluble matter in the mobile phase or sample solution. 2. Dirt in the tubing. 3. Plunger seal fragment. 4. Precipitation of sample components. |
• Sonicate the filter or replace it. • Filtrate the mobile phase and sample solution in advance using a membrane filter. • Attach a line filter. • Clean the tubing and replace the plunger seal. • Prepare a sample solution with the mobile phase. |
2. Peak splitting, tailing, and broadening. |
1. Void in the column head. 2. Dead volume due to inappropriate connections. 3. Inappropriate mobile phase conditions. • Ion suppression method: Inadequate suppression (Too much sample). • Ion-pair method: Inadequate concentration of the ion-pair agent (Too much sample). 4. Column deterioration. * Not repairable in the case of column deterioration or damage to the packing condition. |
• Replenish the packing material. • Reconnect the tubing. • Review the pH, salt concentration, sample amount, and other conditions. • Review the ion pair agent concentration, pH, sample amount, and other conditions. • Check the column performance using standard inspection solution. |
3. Retention time too long or unstable. |
1. Liquid leak (Indicated on the pressure gage of the pump). 2. Inappropriate mobile phase conditions. 3. Inadequate column equilibration time. |
• Check the pump and tubing for any leaks. See 2-3. • Secure adequate equilibration time. |
4. Retention time too short. | 1. Hydrolysis (deterioration) of a bonded groups by strong acid or base. 2. Inappropriate mobile phase conditions. 3. Inadequate column equilibration time. |
See 2-3. • Secure adequate equilibration time. |
SUPERIOREXII ODS is shipped after a strict performance check. However, if you should find any defect, please contact your dealer or Osaka Soda for replacement. Note that Osaka Soda does not warrant the product against column life or deterioration
caused by the failure to follow the above handling instructions.
Ten or more days after reception by the customer, Osaka Soda will assume that the product was delivered in good condition, and will not accept a later replacement request.
Update June 1, 2022