Thank you for purchasing CAPCELL CORE ADME high performance core-shell type columns.
CAPCELL CORE ADME columns are packed with high-performance core-shell type packing material of 2.7-µm particle (consisting of a 1.7- µm solid silica core surrounded by a 0.5-µm surface layer of porous silica), which is developed with combination of technology for silica particle by coating with a single thin layer of silicone polymer and technology for Adamantyl group bonding.
To ensure CAPCELL CORE PFP column stability and long life, please read and follow this manual carefully for the correct of use products.
With CAPCELL CORE, because a core-shell structure substrate has been adopted, the specific surface area of packing material is smaller compared to totally porous material. Consequently, if you intend to change from a totally porous packing material column to a CAPCELL CORE product, and if you are used to injecting samples at the upper load limit, take note that it is necessary to reduce the sample loading volume.
For tubing, follow the instructions in Fig. 1. Inappropriate tubing, particularly when a tube used for a different type of column is used as is, the tip length beyond the ferrule (V in Fig. 1) may not match the length of the end fitting bore (L in Fig. 1), which leads to troubles.
Where L>V, dead volume occurs and may cause the broadening or tailing of peak, or limit separation performance.
Where L<V, inadequate sealing of the ferrule causes liquid leakage. Therefore, it is recommended that the ferrule be
changed concurrently with the column change.
* After frequent column replacement the male-nut may present a crushed ferrule and cause liquid leakage. In such a case, tightening too hard may cause the head portion of the nut to break off. Therefore, replace the ferrule in good time.
Fig. 1
Since it is impossible to enumerate all of the troubles associated with HPLC analysis work have diverse causes, only those which occur relatively often in and around the column are cited:
Symptom | Causes | Remedies |
---|---|---|
1. Rise in column pressure | Clogging with foreign matter 1.Foreign particles, insolubles in eluants and sample solutions 2. Water deposits in the tubing 3. Fragments of plunger seal 4. Precipitation of sample constituents |
• Clean the filter with an ultrasonic device or replace. • Filter the solvent and sample solution through a membrane filter beforehand. • Install a liquid line filter. • Clean the tubing and replace the plunger seal • Prepare the sample solution with the mobile phase |
2. Splitting, tailing or broadening of peak | 1. Occurrence of dead volume due to faulty tubing 2. Inappropriate conditions of mobile phase 3. Deterioration of column * Restoration is impossible when deterioration of column or void in packing material bed has occurred. |
• Dismantle the tubing once and reconnect. • Ascertain pH, salt concentration, sample volume, etc. Ascertain ion-pair reagent level, pH and sample solvent, etc. • Ascertain sample solvent and injection volume. • Verify column performance using the standard testing solution shown in the Column Report. |
3. Retention time too long or unstable | 1. Liquid leakage (this can be judged by the movement of pressure gauge on the pump) 2. Inappropriate conditions of the mobile phase 3. Insufficient time for column stabilization |
• Inspect the liquid leakage from pump and tubing system. • Take sufficient time for stabilization. |
4. Retention time too short | 1. Rupture (deterioration) of bonding group due to the use of strong acid or alkali 2. Inappropriate conditions of mobile phase 3. Insufficient time for column stabilization |
• Verify column performance using the standard testing solution. • Take sufficient time for stabilization. |
While CAPCELL CORE PFP columns undergo strict verification of performance before shipping, if any defect is identified, please inform us or our distributor of it.
However, please take note that we are not responsible for matters related
to column service life or problems caused by the failure to follow the above mentioned instructions for care and use. Please note also that we assume the product is assumed to have been delivered in good condition if no claim is made within
ten days from receipt by the customer, and requests for replacement submitted thereafter will not be accepted.
October 1st 2013
Latest up date: May 27th 2014