Products
HPLC Column
Ceramospher (Chiral) User's Manual
Overview
Ceramospher (Chiral) User's Manual
1. Handling the Column
2. Attaching the Column
3. Analysis
4. Storing the Column
5. End Fittings
6. Replacement Parts and Repair Items
7. Troubleshooting
Ceramospher Chiral RU-1 and RU-2 are unique columns for chiral separation that are based on spherical ceramics.
1. Handling the Column
1. Handle the column with great care. A strong shock may cause damage.
2. Attach or detach the column when the pressure gage indicates zero.
PAGE TOP
2. Attaching the Column
1. The column joint is of the male nut type for tubing of 1/16 inch OD. Check that the piping joints of the system fit correctly and that the ferrule tips are deeply inserted into the joints. (See Fig. 1.)
2. Before attaching the column, replace the liquid in the system with the mobile phase to be used.
(Refer to the section on mobile phases.)
3. Attach the column according to the direction of the arrow.
PAGE TOP
3. Analysis
3-1. Mobile Phase
1. RU-1
1. In addition to methanol, non-aqueous solvents such as ethanol, isopropanol, acetonitrile, chloroform, and n-hexane are acceptable. It is advised to try methanol at first, and to use other solvents only when methanol does not provide a good separation.
2. A mobile phase prepared by adding isopropylamine, diethylamine, triethylamine, or acetic acid of up to 1vol% to methanol is also acceptable.
3. After degassing completely, filtrate the mobile phase using a membrane filter 0.45 µí or smaller to remove dust. A 2-µí filter is used at the column inlet. To prevent foreign matter from clogging the column inlet filter, we recommend using a line filter.
4. Methanol is sealed in the column.
5. When using a mobile phase from the same resevoir over a long period, it is necessary to take chemical or physical measures to remove moisture out of the liquid.
6. To prevent column deterioration, avoid the following:
. Injection of a mobile phase or a sample solution containing water
. Frequent change of the mobile phase composition or direct change to a mobile phase of low compatibility
. Rapid change in column inlet pressure
. High column pressure due to the use of a high-viscosity mobile phase
7. The maximum operating pressure of this column is 20 MPa. To ensure that this pressure is not exceeded, set the flow rate as follows. Set the flow rate and pressure using the following values as the standards (reference values).
(50°C, methanol, and MPa)
0.5mL/min 1.0mL/min 1.5mL/mina 2.0mL/min
4.6mmx150mm 3 6 9 9
4.6mmx250mm 5 10 15 25
(50°C, methanol, and MPa)
2.RU-2
1. Preparing a Mobile Phase
Non-Aqueous Type Aqueous Type (Water Content: 10-70vol%)
Neutral Compound MeOH or acetonitrile Potassium dihydrogenphosphate of 0.5 to 5 mmol/L (No precipitation allowed)
Basic Compound 0.1-1vol% diethylamine / MeOH or acetonitrile 0.01-0.1vol% diethylamine/MeOH or acetonitrile
Acidic Compound 0.1-1vol% acetic acid / MeOH or acetonitrile Acetic acid of 0.01-0.11vol% / MeOH or acetonitrile
potassium dihydrogenphosphate of 0.5 to 5 mmol/L (No precipitation allowed)
Amphoteric Compound Derivatize chemically, and treat as the neutral (above)
*When an acetic acid solution is used, the waste liquid is slightly colored. However, this will not affect column performance. Dispose of the waste liquid as heavy metal-containing liquid.
2. Methanol is sealed in the column at the time of shipment.
3. To prevent column deterioration, avoid the following:
. Frequent change of the mobile phase composition or direct change to a mobile phase of low compatibility
. Rapid change in column inlet pressure
. High column pressure due to the use of a high-viscosity mobile phase
4. The maximum operating pressure of this column is 20 MPa. To ensure that this pressure is not exceeded, set the flow rate as follows:
Methanol 1.0 mL/min Water/Methanol 0.5mL/min
Acetonitrile 1.0mL/min Water/Acetonitrile 0.5mL/min
5. Since potassium dihydrogenphosphate does not easily dissolve in a solution of water/methanol or water/acetonitrile, prepare the solution by the following procedure:
1. Dissolve potassium dihydrogenphosphate completely in water.
2. Add methanol or acetonitrile to the solution while stirring.
3. After degassing, check that potassium dihydrogenphosphate is not precipitated.
3-2. Preparing a Sample Solution
1. Dissolve the sample in a solvent of the same composition as the mobile phase wherever possible.
2. Using a solvent with strong dissolving power may lower separation efficiency or cause the sample to precipitate at the column head.
3. If there is insoluble matter remaining in the sample solution, filtrate the solution using a filter 0.45 µm or smaller.
3-3. Temperature
1.RU-1
Ceramospher RU-1 is extremely temperature stable. Although the operating temperature differs depending on the viscosity and boiling point of the mobile phase, Ceramospher RU-1 can be used in the range from 20 to 50°C if the mobile phase is of the methanol type.
2.RU-2
Ceramospher RU-2 is extremely temperature stable. Although the operating temperature differs depending on the viscosity and boiling point of the mobile phase, Ceramospher RU-2 can be used in the range from 15 to 60°C if the mobile phase is of the water or methanol type.
PAGE TOP
4. Storing the Column
1. Seal the column with the accessory plug and store it in a cold place where there is little temperature fluctuation.
2. After using an aqueous mobile phase or the above mobile phase with acetic acid or amines added, rinse the column with methanol and store it after sealing with methanol.
PAGE TOP
5. End Fittings
1. An analytical column of up to 6-mm ID uses a filter-embedded end fitting as shown in Fig.. The filter cannot be changed alone. If the filter is clogged or the column pressure is high, replace the end fitting. See Table 4 for the replacement parts and repair items.
2. See Fig. 1 for the column connection. If the tubing is inappropriate, especially if a tube for a different type of column is used, the length after the ferrule tip (V in Fig. 1) is often different from the end fitting length L, and a problem may occur.
If L is greater than V, dead volume may be generated and cause peak broadening or tailing or deterioration of separation performance.
If L is smaller than V, liquid may leak because of inadequate ferrule adhesion.
Therefore, we recommend replacing the ferrule together with the column.
*If the column is replaced frequently, the male nut may crush the ferrule and liquid may leak. Since tightening the nut too much may cause its head to come off, replace the ferrule at an early stage.
Fig. 1  Column connection
Fig. 1 Column connection
PAGE TOP
6. Replacement Parts and Repair Items
Table 4 Replacement parts and repair items
Part No. Part Name Description
EF2052 End fitting (4.6 mm) 2 pieces
EF2160 Piping kit(2pcs) Male nuts (1/6) and ferrules (1/16) 2pieces each
EF2161 Ferrule (1/16) Ferrules (1/16)10pieces
PAGE TOP
7. Troubleshooting
Problems in high performance liquid chromatography are attributable to various causes that cannot all be enumerated. The table below describes some comparatively common problems related to the column.
Symptom Cause Measures
1. Column pressure rise.
Blocking with foreign matter
1. Dust or insoluble matter in the mobile phase or sample solution.
2. Dirt in the tubing.
3. Plunger seal fragment.
4. Precipitation of sample components.
Sonicate the filter or replace it.
Filtrate the mobile phase and sample solution in advance using a membrane filter.
Attach a line filter.
Clean the tubing and replace the plunger seal.
Prepare a sample solution with the mobile phase.
2. Peak splitting, tailing, and broadening.
1. Dead volume due to inappropriate connections.
2. Column deterioration.
* Not repairable in the case of column deterioration or damage to interior channel.
Reconnect the tubing.
Check the column performance using standard samples.
3. Retention time too long or unstable.
1. Liquid leak (Indicated on the pressure gage of the pump).
2. Mixing of air.
Check the pump and tubing for any leaks.
See 2-3.
Secure adequate equilibration time.
Purge air. (RU-2)
Use a degasser. (RU-2)
Ceramospher Chiral RU-1/RU-2 is shipped after a strict performance check. However, if you should find any defect, please contact your dealer or Osaka Soda for replacement.
Note that Osaka Soda does not warrant the product against column life or deterioration caused by the failure to follow the above handling instructions.
Ten or more days after reception by the customer, Osaka Soda will assume that the product was delivered in good condition, and will not accept a later replacement request.
Page top