|
A11 |
In this kind of case, a liquid flow usually restores the column. Start the liquid flow as follows:
1. Remove the outlet of the column.
2. Let through the liquid of the initial concentration (initial composition) of gradient. Check that the mobile phase comes out for the column outlet.
*If the flow of the initial composition is not possible, let through acetonitrile of 50% or less. If the aqueous mobile phase is buffer solution and causes precipitation in the column, let through water (if necessary, water of about 40°C). If a high pressure is imposed, reduce the flow rate.
3. Once the liquid flow has recovered, stop pumping once, connect the column outlet, and start pumping again.
4. After checking that the liquid flows to the last outlet of the tubing, set the analytical conditions to the highest acetonitrile concentration and purge air thoroughly.
*Since air may be trapped also in the detector cell, purge air thoroughly.
5. Check that the baseline has become stable and then restart the analysis. |
|